Description
The Delta (DT) Series spectrometers are designed for high-resolution, high-sensitivity optical coherence tomography (OCT) applications. Utilizing a fully transmissive Czerny–Turner optical design (T–T–T configuration) with a volume phase holographic (VPH) transmission grating, the DT Series achieves ultra-high optical resolution (0.04 nm) and exceptional stability. A patented alignment mechanism ensures precise, repeatable calibration. Models are available with 20 kHz or 80 kHz CMOS line-scan cameras, providing flexibility for diverse OCT system architectures. The Delta Series integrates seamlessly via USB 3.0 and supports comprehensive SDK tools for software development and wavelength calibration.
Delta OCT Series Spectrometer
Specifications
| Wavelength Range (reflectance): | 800 – 880 nm |
|---|---|
| Resolution: | 0.04 nm |
| Dimensions: | 180 × 120 × 63 mm (spectrometer only); 60 × 60 × 31.1 mm (camera) |
| Detector Type: | 2048-pixel CMOS line-scan sensor |
| Cooling: | Uncooled (ambient operation, 5–40 °C) |
| Signal-To-Noise Ratio (SNR): | 51 dB (single acquisition) |
| Thermal Stability: | Factory-calibrated alignment repeatability ensured by patented mechanism |
| A/D Resolution: | 10 / 11 / 12-bit selectable |
| Integration / Scan Time Range: | 12.5 µs – 50 µs (model dependent) |
| Minimum Scan Time: | 0.0000125 s |
| Average Minimum Scan Time: | 0.00003125 s |
| Dynamic Range: | 69 dB |
| Stray Light: | < 0.2 % (design-optimized) |
| Interfaces: | USB 3.0 high-speed data interface |
| Connector Type: | 5 µm single-mode FC/PC fiber input |
| Power Requirements: | 6–15 V DC (< 3.6 W, USB 3.0 compliant) |
| Operating Temperature: | 5 °C – 40 °C |
| Storage Temperature: | –30 °C – 70 °C |
| Environmental Limits: | 0 – 85 % RH, non-condensing |
| A/D Converter Type: | 10 / 11 / 12-bit selectable ADC |
Got questions about specs? Use the inquiry form to ask.
Features
- Ultra-high optical resolution: 0.04 nm
- Transmission grating and fully transmissive Czerny–Turner (T–T–T) design
- f/# 3.6, NA 0.14; focal length 60–89 mm @ 840 nm
- Wavelength range: 800–880 nm
- Available with 20 kHz or 80 kHz CMOS line-scan cameras
- High sensitivity and low dispersion for OCT performance
- 5 µm single-mode FC/PC fiber input
- USB 3.0 high-speed data interface
- Supports wavelength calibration and SDK-based software integration
- Designed for ophthalmic and tissue OCT systems
Applications
- Optical Coherence Tomography (OCT) Systems
The DelTaOCT Spectrometer is engineered for integration into advanced OCT systems, enabling ultra-high-resolution cross-sectional imaging for research and clinical applications. - Ophthalmic Imaging
Ideal for retinal imaging, glaucoma assessment, and other ophthalmic diagnostics, providing precise visualization of ocular structures. - Dermatologic and Tissue Imaging
Supports non-invasive imaging of skin and soft tissues, facilitating early detection of dermatological conditions and monitoring of tissue health. - Biomedical Diagnostics
Enables high-sensitivity spectral analysis for early disease detection, cellular imaging, and pathology research in medical laboratories. - Industrial Metrology and Non-Destructive Testing
Used for high-precision thickness measurements, layer inspection, and flaw detection in materials and components without causing damage. - Semiconductor and Surface Inspection
Provides detailed surface profiling and defect analysis for semiconductor wafers, microelectronics, and precision optics manufacturing. - Material Science Research
Facilitates the study of material properties, microstructure analysis, and characterization of composites and polymers. - Pharmaceutical Quality Control
Used in the inspection of pharmaceutical products, ensuring consistency and detecting impurities or structural anomalies. - Custom Spectroscopic Applications
With SDK and USB 3.0 support, the DelTaOCT Spectrometer can be integrated into custom-built imaging and measurement platforms for specialized research and industrial needs. - Academic and Clinical Research
Supports a wide range of scientific investigations in universities, hospitals, and research institutes, including optical imaging, tissue engineering, and biomedical optics.
Frequently Asked Questions
What is the Delta Spectrometer designed for?
What optical design does the Delta Spectrometer use?
What are the available models and their specifications?
What is the optical resolution of the Delta Spectrometer?
How does the Delta Spectrometer connect to other systems?
What are the power requirements for the Delta Spectrometer?
What are the environmental operating conditions for the Delta Spectrometer?
Got more questions? Use the inquiry form to ask the supplier directly.
Similar Products
Your inquiry has been received.
Create an account by adding a password
Why create an account?
- Auto-complete inquiry forms
- View and manage all your past messages
- Save products to your favorites
- Close your account anytime — no hassle