Description

FLD-KIT is a set with everything you need to quickly create a suitable power supply for a lamp-pumped laser. Develop and launch your laser (e.g. Nd:YAG, Er:YAG, CTH:YAG, Alexandrite) or IPL system easy with a flexible set of our modules.

The resulting power supply is able to charge a capacitor bank up to defined value, provides initiating flash voltage pulses with required width and repetition rate (in internal synchronization mode) or with required width (in external synchronization mode), maintains a simmer current during the interval between pulses.

We have composed some kits suitable for standard applications, our customers often asked for.

  • Holmium lasers for surgery (Ho:YAG, CTH:YAG) with 1, 2 or 4 pump chambers
  • Alexandrite, long pulse Nd:YAG, long pulse KTP laser systems for aesthetics for free-discharge and PWM-based systems
  • Nanosecond (q-switched) and Picosecond Nd:YAG laser power supply, systems for tatoo removal (Q-switched Nd:YAG, picosecond Nd:YAG)
  • Er:YAG laser system for dermatology
  • IPL machines (for example, high power free-discharge based system, dual-channel (PWM and free discharge) system with the possibility of truly rectangular long pulses)

Multifunctional machines

  •   IPL / laser / RF machines with handheld applicators (low-power low-voltage)
  •   IPL / laser / diode machine with handheld applicators (medium power)

 

Modules:

  • 1 kW capacitor charger PCA-10 , 2 kW capacitor charger PCA-20 or 4 kW capacitor charger PCA-40
  • Discharge circuit, e.g. NBU-1012 or other
  • Control board LSCB + Software
  • Capacitor bank module
  • Other modules like Pockels cell driver QBU,  Laser diode driver PDD-1000 or RF Generator RFGM (ask us for the details) 

 

FLD-kit Flashlamp Driver set from OEM Tech modules

Specifications

Output Current (max): 3000 A
Output Voltage (max): 1500 V
Average Output Power: 8000 W
Number Of Driving Channels: 1 – 10+
Output Voltage: up to 1500 V
Output Current: up to 3000 A
Output Power: up to 8 kW
Discharge Circuits: Thyristor- or IGBT-based

Got questions about specs? Use the inquiry form to ask.

Document icon Download Data Sheet Download icon

Features

  • Holmium Lasers for Surgery (Ho:YAG, CTH:YAG)
    • 1 Pump Chamber (25-30W Optical Power)
      • PCA-20-800V-PD - 1pc (Suggested voltage: 800V; 900V and 1000V versions also available)
      • NBU-1012 - 1pc
      • CB-900V-3.5mF - 1pc
      • LSCB - 1pc (Standard version)
    • 2 Pump Chambers (50-60W Optical Power)
      • PCA-40-1000V-PD - 1pc (Suggested voltage: 1000V; 800V and 900V versions also available)
      • NBU-1012 - 2pcs
      • CB-900V-3.5mF - 1pc (Custom 1000V capacitor bank available if voltage exceeds 900V)
      • LSCB - 1pc (Standard version)
    • 4 Pump Chambers (100-120W Optical Power)
      • PCA-40-1000V-PD - 2pcs (Connected in parallel for 8kW power; Suggested voltage: 1000V; 800V and 900V versions also available)
      • NBU-1012 - 4pcs
      • CB-900V-3.5mF - 2pcs (Custom 1000V capacitor bank available if voltage exceeds 900V)
      • LSCB - 1pc (4 channel version)
  • Alexandrite, Long Pulse Nd:YAG, Long Pulse KTP Laser Systems for Aesthetics
    • Free-Discharge Based System
      • PCA-40-700V-PD - 1pc (Suggested voltage: 700V; other versions available)
      • NBU-1012 - 2pcs (One NBU per flashlamp)
      • CB-350V-20mF - 8pcs (Customer can rearrange cap. banks)
      • LSCB - 1pc (Standard version)

Applications

Power supply for a lamp-pumped laser / IPL / RF / Diode systems

Frequently Asked Questions

What types of laser systems are available in the datasheet?
The datasheet includes Holmium lasers for surgery, Alexandrite, long pulse Nd:YAG, long pulse KTP laser systems for aesthetics, Er:YAG laser systems for dermatology, Nanosecond (q-switched) and Picosecond Nd:YAG laser systems for tattoo removal, and IPL machines.
What is the suggested voltage for the PCA-40-1000V-PD in the 2 pump chambers configuration?
The suggested voltage for the PCA-40-1000V-PD in the 2 pump chambers configuration is 1000V, although 800V and 900V versions also suit well to the application.
Can the capacitor banks be customized for different voltage requirements?
Yes, if the operating voltage exceeds 900V, a custom 1000V capacitor bank can be requested.
What is the purpose of the QBU-5020 component in the Q-switched Nd:YAG laser system?
The QBU-5020 is a Q-switch driver used in the Q-switched Nd:YAG laser system.
What is the function of the LSCB component in these systems?
The LSCB component is a control board, which provides parameters control and coordinated operation of power modules by OEM Tech to drive laser or IPL system flashlamp(s). It can vary slightly between systems, such as 'PWM' version for PWM-based systems or a 'pico' version for Picosecond Nd:YAG laser systems.
What is the recommended configuration for achieving 8kW power in a 4 pump chamber setup?
To achieve 8kW power in a 4 pump chamber setup, two PCA-40-1000V-PD components should be connected in parallel.

Got more questions? Use the inquiry form to ask the supplier directly.

Similar Products

Request Quote